The purpose of this study is to assess the diagnostic accuracy of the Siemens Biograph mCT 64 scan TrueV UHD CT scan, which in reality is a double 32-slice CT scan, in patients living in the North East of the Netherlands suspected of coronary artery disease (CAD). The findings of the computed tomography coronary angiography (CTA) are compared to the findings of the golden standard, in this case invasive coronary angiography (ICA) with eventual an instantaneous wavefree ratio (iFR) or an fractional flow reserve (FFR) measurement. 662 patients underwent CTA, in which 552 patients didn’t underwent ICA. 53 patients were excluded for several reasons. 77 patients underwent CTA and ICA. The diagnostic accuracy of the CTA to prove or rule out significant CAD (>50% stenosis) is investigated per patient, per segment and per whole vessel. For patient-based analysis, CTA had a sensitivity of 64,5%, a specificity of 60,9%, a positive predictive value of 52,6% and an negative predictive value of 71,8%. For the segments, CTA had a sensitivity of 37,2, a specificity of 95,8%, a positive predictive value of 45,5% and an negative predictive value of 94,5%. For the whole vessel, CTA had a sensitivity of 50,9%, a specificity of 88,9%, a positive predictive value of 44,5% and an negative predictive value of 90,3%. Our results demonstrates that the Siemens Biograph mCT 64 TrueV UHD CT scan is just as reliable to rule out significant CAD compared to a 64-slice CTA. And therefore, it could be used as a safe alternative instead of the ICA in patients with low to intermediate risk of CAD. Because of the low numbers of true positive results on the CTA there can been none pronouncement done regarding the ability to prove CAD.

Item Type: Thesis (Thesis)
Supervisor name: Facultair en tweede begeleider and Anthonio, Dr. R.L.- cardioloog Scheper Ziekenhuis Emmen
Supervisor name: Schuurman Drs. R., cardioloog Scheper Ziekenhuis Emmen and Bie, Dr. M. De Refaja Ziekenhuis Stadskanaal
Faculty: Medical Sciences
Date Deposited: 25 Jun 2020 10:53
Last Modified: 25 Jun 2020 10:53